DEFIBRILLATION ELECTRIC SHOCK
Waveform: Biphasic Truncated Exponential (BTE) waveform (Impedance compensation)
Energy:
Adult: 185 to 200J (±5%)
Paediatric: 45 to 50J (±5%)
Operating Mode: Semi-Auto
Charge Time: Less than 12 seconds
Analysis Time: 8-13 seconds
Shock Time: Delivers 200 shocks with full battery
ECG
Lead: II (RA, LL)
Patient Impedance: 25 to 175ohm
Detection:
V/Fmore than 200µV
V/Tmore than 160 per min
INDICATION
Controls
Standard:
Open (power) switch, shock button,
Adult/Paediatric mode switch
Indicators
Visible:
ICON Indicator, Status LCD (unit status, battery status, temperature status),
LED (Adult/Paediatric mode switch LED)
Audible:
Speaker (voice prompt),
Beep (CPR indication)
PHYSICAL
Dimensions: 294H x 240W x 95mmD
Weight: Approx. 2.65kg (including battery, excluding pads)
SELF TEST
Cycle: Every 24 hours, 1 week, 1 month Power on self test, battery insertion self test
Test Result: Status LCD displays ‘O’ / ‘X’
DATA BACKUP AND COMMUNICATION
Standard SD card,
IR communication port
ACCESSORY SPECIFICATIONS
Adult/Paediatric Pads
Standby Life: 2 years from date of manufacture
Electrodes: Disposable pads
Placement: Adult: Anterior-lateral
Paediatric: Anterior-posterior
Minimum active gel area: 80cm2 +/-5%
Cable Length: Approx. 1.8m
Environmental Conditions
Temperature: Operation: 0 to 43°C (32 to 109.4°F)
Storage: 0 to 43°C (32 to 109.4°F)
Relative Humidity: 5 to 95% RH (non-condensing)
BATTERY
Type: LiMnO2, disposable, long-life primary cell
Voltage/Capacity: 15V, 4200 mAh
Shelf Life: 2 years from date of manufacture (in the original packaging)
Standby Life: 5 years from date of manufacture (inserted in the AED)
Discharge: A minimum of 200 shocks (except the CPR period between the defibrillation therapy) or 10 hours of operating time at 20°C
Environmental Conditions
Temperature: Operation: 0 to 43°C (32 to 109.4°F)
Storage: 0 to 43°C (32 to 109.4°F)
Relative Humidity: 5 to 95% RH (non-condensing)
Information correct at time of publication. Subject to change without notice.